ExpeData Specifications Print
Overview   Specifications   Details  Gallery  Brochure   Add to RFQ 
ExpeData Specifications:
Number of Channels: Up to 96 during data acquisition; unlimited (up to RAM limit) during data analysis
Computer Required: Desktop or laptop computer running Windows 98 or above
Standard Data Acquisition Interface: Sable Systems UI-2 (see below for specifications)
Other Supported Acquisition Interfaces: Direct parsing of numerical data from the clipboard; direct reading of data from a virtual hard drive shared by other applications, e.g. VideoDaemon ; all can be used simultaneously
Number of Samples: Unlimited (up to RAM limit)
Real-time Graphic Display: Standard
Interface: Fully menu-driven (no programming required)
Acquire Data in User Units: Fully supported, including non-linear transforms
Panning and Zooming During Analysis: Standard, including mouse-wheel zoom
Macro Support: Fully supported: Record, playback, pause; automate transformation and analysis; scripts are created during macro recording and are fully editable 
Direct Writing of Statistics to Spreadsheet: Fully supported, under manual or macro control
Internal Spreadsheet: Supports direct writing from macros; single-click transfer to Excel; very versatile transformation script utility
Internal Statistics Package: For spreadsheet: Supports missing data; Means, SDs, skewness, kurtosis, linear regressions, multiple regressions, polynomial regressions, ANOVA, ANCOVA, t and Mann-Whitney U tests and more
UI-2 Technology: Successive approximation analog to digital conversion (16 bits)
UI-2 Analog Input Ranges: -5.12V to +5.12V (8 main voltage inputs), -5 to 60.5 C (4 accessory inputs in temperature measurement mode), 0 to 5.12V (4 accessory inputs in voltage measurement mode), 10 ohms to 10 megohms (4 accessory inputs in resistance measurement mode)
UI-2 Analog Outputs: Two, 12 bits resolution, 0 - 5V
UI-2 Counter: One, 16 bits (rollover at 65536 counts)
UI-2 Digital Outputs: Eight, TTL level (off = 0V, on = 5V), short circuit proof
UI-2 Sampling Rate: 100 Hz/channel in normal operating mode; up to 1 kHz/channel with internal digital filtration turned off
UI-2 Interface: RS-232 serial, 115200 baud (works with any USB to serial converter)
UI-2 Protocol: Fully open; available on request
UI-2 Power Requirements: 9 - 18VDC (12 VDC nominal), < 20 mA. 110VAC or 220VAC adapter supplied
UI-2 Operating Temperature: 0 - 60 degrees C, relative humidity non-condensing
UI-2 Size/Weight: 7" x 6" x 3" (17 x 15 x 7 cm); less than 2 lbs (1 kg)
 