| Oxzilla II Specifications: |
| Technology: Dual fuel cell, temperature controlled, not harmed by moisture, CO2 or particulates; 18-36 month life, user-replaceable (10 minutes). |
| Range: <1 - 100% O2, each channel |
| Accuracy and Resolution: Accuracy 0.1% (1-100%), resolution 0.0001% |
| Noise and Drift: At constant oxygen concentration and recommended 5-second filtration, noise < 8 ppm pk-pk typical, drift < 0.02%/24 hrs |
Response time: <10 seconds, 10% to 90% of a step change
|
Flow Rate and Moisture: 10 - 1000 ml/minute, relatively insensitiveto flow rate variations
|
| Barometric Pressure Compensation: Selectable; barometric pressure resolution 0.0001 kPa |
| Four Analog Outputs: Channels 1 and 2, delta (1-2), and barometric pressure; multiple ranges for each, 0-5V 16 bit outputs, optically isolated |
| Serial Data Output: Absolute O2 concentration or partial pressure on channels 1 and 2, delta (1-2), all to 0.0001%; barometric pressure to 0.0001 kPa; cell temperature to 0.001 C, comma delimited ASCII, at user-set period of 0.2 - 1000 sec |
| Digital Filtration: Settable from zero to 40 sec in 0.2 sec increments for 0 to 63% of a step change |
| Internal A/D Conversion: 24 bits |
| User Interface: Intuitive, menu-based system with non-volatile storage of settings |
| Panel Display: Line 1: Channel 1 O2 to 0.0001%. Line 2: Channel 2 O2 to 0.0001%. Line 3: Delta O2 to 0.0001%. Line 4: Barometric pressure to 0.0001 kPa |
| Temperature Compensation: Automatic, controlled temperature settable from 10 to 50 C in 1 C increments or turned off |
| Tubing Fittings: Nickel-plated barb fittings for 1/8" or 3 mm i.d. semi-rigid tubing; other fittings available on request |
| Power Requirements: Temperature regulation on: < 100W. Temperature regulation off: < 5W |
| Operating Temperature: 5 - 45 C, non-condensing humidity |
| Size/Weight: [fill in BJ] |
| |