| PA-10a Specifications: |
Technology: Paramagnetic sensor (never needs replacement)
|
| Range: 0% to 100% O2, 0 to 105+ kPa O2 |
| Accuracy and Resolution: At constant temperature and flow rate, accuracy better than 0.1%, 0 - 100% O2; Resolution 0.0001% |
| Gas Ports: 1/8" barb fittings on front panel, input line filtered |
| Flow Rates: At constant temperature, accuracy better than 0.1% of full scale, 1 - 100% oxygen, resolution 0.0001%O2; Barometric pressure accuracy better than 0.05% of full scale, resolution 0.0001 kPa |
| Noise and Drift: At constant temperature, flow rate and O2 concentration, 2 second filtration: Noise < 0.002% pk-pk over 20 minutes; Drift < 0.02% over 24 hours. (measured at atmospheric oxygen concentrations) |
| Panel Display: Digital 2-line alphanumeric, simultaneous display of oxygen (0.0001 % or 0.0001 kPa resolution) and barometric pressure. (0.0001 kPa resolution) |
| Voltage Outputs: Oxygen: 0 - 5 Volts for 0 - 25, 19 - 21.5, 0 - 100, 0 - 125, and 80 - 105 %/kPa. Other ranges available to order. Barometric pressure and sensor temperature, four ranges each, 0 - 5 volts. All outputs short circuit protected; 16 bit (up to 1 in 65535) resolution in each range |
| Serial Output: Percent oxygen, kPa oxygen (both to 0.0001%/kPa), barometric pressure (to 0.0001 kPa), and internal analyzer temperature (to 0.001 deg C) with comma delimiters; resolution 1 part in 1,000,000 for concentration and pressure; baud rate adjustable from 9600 to 115200 in 5 steps |
| Power Requirements: 8 - 18 VDC at approx. 5 W, reverse polarity protected. (UL/CSE 120 or 220 VAC AC adapter supplied with unit) |
| Size: 13" x 10" x 4" (33 x 25 x 10 cm) |
| Weight: Less than 10 lb (4.5 kg) |
| |
|