SA-4 Specifications:
Technology: Oxygen analysis: Fuel cell; CO2 analysis (optional): Non-dispersive dual wavelength infrared; Water vapor analysis (optional): Bulk polymer capacitive sensor; Mass flow measurement and control: Thermal mass flow sensor
Body Size Range: Typically large arthropod to rat in flow-through mode, depending on flow rate used and other factors
Flow Rate: 25 - 1500 ml/minute
Flow Regulation: PID (Proportional Integral Derivative) controlling pump directly for minimum pump wear and noise
Oxygen Analyzer: Sable Systems FoxBox
Carbon Dioxide Analyzer: Optional; Sable Systems FoxBox (integrated O2 and CO2 analyzer)
Water Vapor Analyzer: Optional; Sable Systems RH-300; or user-supplied analyzer
Data Acquisition: Sable Systems UI-2 interface (optional); ExpeData software standard
Non-Volatile Data Storage: 8000 points, each comprising all channels of data; independent of external data acquisition; inter-point interval adjustable from 0.1 sec - 900 sec; multiple recordings supported; upload utility included
Data Analysis: Sable Systems ExpeData with full macro capability for automated analysis if desired
Technical Support: Free, lifetime
Warranty: Three years
Setup and Training: Optional, generally not required for experienced investigators
Power Requirements: Nominal 12 VDC; Universal adapter supplied
Operating Temperature: 0 - 45 degrees C, relative humidity non-condensing
 