| UI-2 Specifications: |
| Technology: Successive approximation analog to digital conversion (16 bits) |
| Analog Input Ranges: -5.12V to +5.12V (8 main voltage inputs), -5 to 60.5 C (4 accessory inputs in temperature measurement mode), 0 to 5.12V (4 accessory inputs in voltage measurement mode), 10 ohms to 10 megohms (4 accessory inputs in resistance measurement mode) |
Analog Outputs: Two, 12 bits resolution, 0 - 5V
|
Counter: One, 16 bits (rollover at 65536 counts)
|
Digital Outputs: Eight, TTL level (off = 0V, on = 5V), short circuit proof
|
Sampling Rate: 100 Hz/channel in normal operating mode; up to 1 kHz/channel with internal digital filtration turned off
|
Interface: RS-232 serial, 115200 baud (works with any USB to serial converter)
|
| Protocol: Fully open; available on request |
| Power Requirements: 9 - 18VDC (12 VDC nominal), < 20 mA. 110VAC or 220VAC adapter supplied |
| Operating Temperature: 0 - 60 degrees C, relative humidity non-condensing |
| UI-2 Size/Weight: 7" x 6" x 3" (17 x 15 x 7 cm); less than 2 lbs (1 kg) |
| |
|