Specifications:
Accuracy and Resolution:Dewpoint accuracy 0.25 C, resolution 0.01 C
Analog Output:Multi-range for relative humidity, dewpoint or water vapor pressure; 0-5V output with 16 bit resolution
Condensate removal:Miniature peristaltic pump, user-settable interval 1-30 minutes, actuation duration user-settable from 1 sec to 30 sec
CondensateRemoval:Miniature peristaltic pump, user-settable interval 1-30 minutes, actuation duration user-settable from 1 sec to 30 sec
Dewpoint Range:1-45 degrees C, in 0.1 C increments
Flow Rate:25 ml/minute to 3 liters/minute
Operating Temperature:5 - 35 degrees C, relative humidity non-condensing
Panel Display:Digital 2-line alphanumeric LCD, simultaneous display of set and read values
Power Requirements:Nominal 12 VDC at 5A; external 90-240 VAC input power supply inlcuded with unit
Regulation::PID (Proportional Integral Derivative), dewpoint deviation < 0.05 C RMS typical
Relative Humidity Range:At 25 degrees C target temperature, approximately 20 - 95+ RH% attainable; settable from 0-100%RH in 0.1%RH increments
Serial Data Output:Comma delimited ASCII string of all measured parameters, standard RS-232 serial, polled (for remote control of setpoint) or auto-sent at user definable intervals from 0.5 - 600 seconds.
Size/Weight:8" x 9" x 10"(19 x21 x26 cm); 13 lbs (6 kg)
Technology:Solid state heat pump (Peltier effect)
Tubing Fittings:Nickel plated barb fittings for 1/8" (3 mm) i.d. flexible or semi-rigid tubing; other fittings available to special order
Water Vapor Pressure Range:0.65 kPa - 9.6 kPa, in 0.01 kPa increments

Click on image to close