Specifications:

Accuracy and Resolution: Accuracy 0.25 C, resolution 0.01 C
Analog Output: 0-5 C = 0-5V output with 16 bit resolution
Condensate removal: Miniature diaphragm pump, user-settable actuation from 1/minute to 1/30 minutes, actuation duration user-settable from 1 sec to 30 sec
Flow Rate: 25 ml/minute to 5 liters/minute
Internal A/D Conversion: 16 bits
Operating Temperature: 5 - 35 degrees C, relative humidity non-condensing.
Panel Display: Digital 2-line alphanumeric LCD, simultaneous display of set and read temperature
Power Requirements:12 VDC at 8A. Universal (90-240 VAC) external power supply included.
Range: 1-5 degrees C, in 1 C increments
Regulation: PID (Proportional Integral Derivative), deviation < 0.05 C RMS
Serial Data Output: Comma delimited ASCII string of all measured parameters, resolution 0.01 C, standard RS-232 serial, polled or auto-sent at user definable intervals from 0.5 - 600 seconds.
Size/Weight:8" x 9" x 10"(19 x21 x26 cm); 13 lbs (6 kg)
Technology: Solid state heat pump (Peltier effect)
Tubing Fittings: Nickel plated barb fittings for 1/8" (3 mm) i.d. flexible or semi-rigid tubing; other fittings available to special order.
User Interface: Intuitive menu system backed by nonvolatile memory of user settings. Single-click spanning at default span settings.