TC-2000 Type-T Thermocouple Meter

Measures 4 channels of temperature using Type T thermocouple probes

TC-2000 Type-T Thermocouple Meter

Measures 4 channels of temperature using Type T thermocouple probes

Features        Description       Specs        Downloads       Contact Us       Sable Enables

Features      Description     Specs      Downloads     Contact Us     Sable Enables

Features

  • -100 to +125C at 0.01C resolution
  • Uses type-T thermocouples
  • Accuracy better than 0.2C, NIST-traceable
  • Multi-range analog plus serial outputs
  • Flexible operation including max/min
  • Rugged, light, low power consumption

DESCRIPTION

The TC-2000 type-T thermocouple meter is a compact, accurate temperature monitoring device. With unparalleled channel correlation and resolution to 0.01°C, the TC-2000 is perfect for differential temperature monitoring applications or temperature monitoring in the field.

This unit offers four channels of temperature measurement using standard, readily available type T thermocouple probes. All four channels are measured five times per second, and all measured temperatures are displayed simultaneously on the easy-to-read display. The TC-2000’s outstanding accuracy (0.2C), resolution (0.01C), and channel agreement (0.02C) set it apart from standard thermocouple meters.

Four channels of matched, multi-range analog outputs plus a standard RS-232 serial output are easily accessed on the back panel. The TC-2000 thermocouple meter can be easily integrated into any data acquisition system.

The accuracy of the TC-2000 thermocouple meter is ensured by a NIST-traceable calibration at our facility. User calibration is easy, using one or more international temperature standards for continued traceability. The TC-2000 thermocouple meter uses an internal reference junction tightly coupled to its reference temperature sensor, making it almost completely insensitive to sudden changes in ambient temperature.

specs

Temperature Measurement  
  • Range:
-100 to 125 deg C
  • Accuracy:
0.2 deg C
  • Resolution:
0.01 deg C
  • Noise:
< 0.01 deg C RMS without filtration
  • Thermocouple Type:
T
Digital Filtration
  • Range:
0 to 50 sec (for 0-63% of a step change)
  • Increment Rate:
0.2 sec
Analog Outputs  
  • Voltage Range:
0-5V
  • Number of Outputs:
1 per channel (total of 4)
Resolution: 16 bits
Connector Type: BNC
Digital Outputs  
  • Communication Standard:
RS232-8N1
  • Baud Rate:
152000
  • Modes:
Broadcast or Polled
  • Sampling Interval:
Configurable from 0.2 to 180 sec
  • A/D Resolution:
16 bits
  • Connector Type:
DB-9
Environmental Requirements  
  • Temperature:
0 – 50 deg C
  • Relative Humidity:
Non-Condensing
Panel Display: Digital, 4 x 16 alphanumeric LCD
Recommended Calibration Interval: 1 Year
Power Requirements  
  • External Power:
12 – 24 VDC; universal adapter supplied
  • Optional Power:
6x AA cells (not included)
Dimensions (W x D x H): 6.3 x 8.0 x 4.0 in. (16 x 20.3 x 10.2 cm)
Weight: 2.2 lb (1 kg)

downloads

Brochure PDF

contact us

Please fill out the form here and our specialists will help configure the right system for your needs or provide you with a detailed quote.

To request manuals and software updates, or find warranty and service information, please contact our Support department.